O-Two CAREvent MRI, Recertified

O-Two CAREvent MRI, Recertified

Item Number: 01CV7000 R
Condition: oxygen accessoriesRecertified
Log in to see your price List price: $6,385.00 EA
1

Description

The O-two CAREvent® ATV+ & MRI are pneumatically powered, time/volume cycled, intermittent positive pressure patient ventilators specifically designed for use by trained personnel in the pre-hospital, hospital, interhospital and air ambulance environments. They do not require batteries or any electrical supply to operate. The various modes of operation of the ventilators support the resuscitation and transportation of a wide range of patients (from infants to adults). The CAREvent® MRI contains no ferrous metals making it suitable for use in an MRI environment up to 3.0 Tesla. The independent controls for Minute Volume and Frequency provided by the CAREvent® ATV+ & MRI offer a comprehensive range of 12 ventilation frequencies and 12 minute volume settings providing 144 possible combinations of delivered tidal volume and respiratory rate. The ergonomically designed and colored control groupings facilitate the selection and setting of the breathing parameters while protecting the controls. The color groupings on the controls add to this “ease of use” concept. The design of the micro-pneumatic circuitry maintains a consistent I:E ratio (set at 1:2) and uses virtually no drive gas. The “Demand Breathing” feature allows spontaneously breathing patients to breathe on oxygen through the ventilator at their own rate and volume. If the demanded rate is sufficient, the “Automatic Circuit Shut Off” feature will cause the automatic cycling of the ventilator to cease when the patient takes an adequate spontaneous breath. The automatic cycling will re-start automatically should the patient stop breathing or his/her demanded volume or respiratory rate drop below the pre-determined minimum.

Compliance with the latest Guidelines for Resuscitation is provided on the CAREvent® ATV+ & MRI by the incorporation of a “Manual Override Button”. This allows for the provision of ventilations and chest compressions during CPR in the patient with an unprotected airway in accordance with these Guidelines. The manual flowrate is equivalent to the flowrate of the minute volume setting selected. For the intubated patient simply set the frequency control to 10 breaths per minute. Unique “audible and visual pneumatic alarms” for circuit disconnection (BSI), low gas supply pressure, low airway pressure and the audible high airway pressure alarm are provided to improve the level of patient safety by alerting the operator to any problems with the patient’s breathing parameters. An override control is provided for the BSI alarm to silence the alarm for 15 seconds. The adjustable “CPAP/PEEP” mode allows for these two functions to be supplied by the ventilator, eliminating the need for an external PEEP valve or CPAP device. To improve ventilator operating time during transport, or when 100% O2 is not required, an “Air Mix Mode” with replaceable environmental filter is provided allowing for an oxygen concentration of either 100 or 60%. The CAREvent® ATV+ & MRI are compatible with air ambulance use, having passed the tests for both fixed and rotary wing aircraft to the US Military Standard 810E.

FEATURES

  • Totally pneumatic operation
  • Durable construction
  • Independent Minute volume and Frequency controls
  • Manual/Auto Selector Switch
  • Automatic & Manual Cycling
  • Low input Pressure Alarm
  • Demand Breathing Capability with auto circuit shut off
  • Adjustable Pressure Relief
  • *G2K Compliant Version Available

SPECIFICATIONS

  • MINUTE VOLUME: 2 – 14 litres
  • BREATHS PER MINUTE: 8 – 40
  • I:E RATIO: 1:2
  • AUTOMATIC FLOW RATE: 12- 40 LPM
  • MANUAL FLOW RATE: As per selected setting
  • DEMAND FLOWRATE: Minimum 120 LPM
  • DEMAND TRIGGERING PRESSURE -5 cm H20 @ 100 LPM
  • MINIMUM INPUT PRESSURE: 45 PSI
  • MAXIMUM INPUT PRESSURE:70 PSI
  • MAXIMUM AIRWAY PRESSURE: 60 cm H2O
  • OXYGEN CONCENTRATION: 60 or 100%
  • ADJUSTABLE AIRWAY PRESSURE RANGE: 20 – 60 cm H2O
  • OPERATING TEMP: -18°C to +50°C (0°F to +122°F)
  • STORAGE TEMP: -40°C to +60°C (-40°F to + 140°F)
  • PATIENT CONNECTION: 9/16” DISS
  • PATIENT VALVE DEADSPACE: 8 ml
  • GAS SUPPLY STATUS INDICATOR: VISUAL
  • LOW INPUT PRESSURE ALARM: PULSED VISUAL and AUDIBLE
  • CYLINDER DURATION: AT 5L MV – 75minutes (approx.)
  • (Based on Aluminum “D” size cylinder containing 415 litres of oxygen)
  • (All specifications are subject to a tolerance of +/- 10%
  • except the I:E Ratio which is subject to a tolerance of +/- 20%)

REFERENCE LIBRARY

Product Brochure

WHAT’S INCLUDED

Includes:

- Oxygen Supply Hose
- Letter of Re-Certification
- Operating Manual
- 12-Month Warranty on Parts & Labor