Baxter Colleague CS3, Recertified
Description
The Baxter Colleague® CS3 Triple Channel Infusion Pump can be utilized for continuous or intermittent delivery through clinically acceptable routes of administration such as, intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces applications. The Baxter Colleague® Pump is offered in four different models; CS (Single Channel w/Monochrome Display), 3CS (Three Channel w/Monochrome Display), CX (Single Channel w/Color Display) and 3CX (Three Channel w/Color Display).
Fluid delivery applications include
- Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
- Whole blood and blood products.
The Colleague Volumetric Infusion Pump feature spectrum includes
- Micro and Macro Rate Range
- Basic Delivery Programming
- Piggybacking Secondary Medications
- Special Programming functions for Dosing
Specifications
- Ac power Requirements: 100-120 VAC, 60Hz
- Leakage Current: Less than 300 u amps earth leakage (tested per UL 2601)
- External Fuses: 1.6 amp Type T (time delay) 250V
- Power Cord: 2.9 m (9 feet) long with integrally molded hospital grade plug
- Battery Supply System: Colleague 3 Pump - 3 hours operating time at 100 mL/hr for all three channels (fully charged new batteries) 12 hours to recharge to 80% of capacity Internal charge system recharges batteries whenever the device is connected to an AC outlet Range of Programmable Flow Rates Primary Infusion on each pump channel 0.1 to 99.9 mL/hr in 0.1 mL/hr increments 1 to 1200 mL/hr in 1 mL/hr increments Piggyback Infusion on each pump channel 0.1. to 99.9 mL/hr in 0.1 mL/hr increments 1 to 500 mL/hr in 1 mL/hr increments Note: Rate limits can be configured for values less Than those stated above.
- Volume to be Infused: 0.1 - 99.9 mL in 0.1 mL increments 1- 9999 mL mL/hr in 1 mL increments. Note: Volume to be Infused limits can be configured for values less than those stated above
- KVO: 0.1 - 5 mL/hr in 0.1 mL increments (configurable option) or programmed rate whichever is less.
- Priming rate: 500 mL/hr (to be used only when not connected to patient)
- Advance Air Rate to Advanced an Air Bubbles: At the same rate programmed for the current primary or piggyback infusion.
- Air Bubble Setting: Measured accumulations of approximately
- Parenteral fluids, drugs and electrolytes; e.g., cardiovascular drugs, antibiotics, anesthetics, analgesics, chemotherapy agents, total parenteral nutrition products, lipids, solutions for irrigation procedures, etc.
- Whole blood and blood products.